Efficacy of a Novel Intraoperative Engineered Sharps Injury Prevention Device: Pilot Usability and Efficacy Trial.
Abstract
[BACKGROUND] The American College of Surgeons reports 88,320 intraoperative needlestick injuries (NSIs) per year, resulting in US $376 to US $2456 in costs per NSI. Engineered sharps injury prevention (ESIP) devices protect against NSIs. To our knowledge, no study has been published to date to demonstrate clinical effectiveness of an intraoperative ESIP device. Operative Armour is a wearable arm cuff that can be donned during surgical closure to allow surgeons to keep a suture pack and sharps protection container on their forearm.
[OBJECTIVE] We characterize Operative Armour's ESIP device effectiveness in a tertiary hospital, hypothesizing that this device will decrease NSI risk by decreasing behaviors associated with NSIs: needle passing and handling.
[METHODS] A prospective case-control study was conducted with institutional review board quality improvement designation in which authors observed skin closures of plastic surgery procedures. To ensure accuracy, one surgeon was observed at a time. Control surgeries were purely observational; intervention cases involved surgeon use of the device during skin closure. Outcomes of interest included needle passing, needle handling, lost needles, and loaded waiting needles.
[RESULTS] Surgeons were observed in 50 control and 50 intervention cases. Operative Armour eliminated needle passing during skin closure. One NSI occurred in one control case; no NSIs were observed in intervention cases (P=.36). The mean number of loaded and unprotected waiting needles was also significantly decreased in the intervention group from 2.3 to 0.2 (P<.001). Furthermore, a multivariable linear regression established that Operative Armour significantly decreased the number of needle adjustments by hand per stitch observed (F=3.72; P=.01). In fact, needle adjustments by hand decreased overall (1 adjustment per 10 stitches vs 1 adjustment per 5 stitches, P=.004), and adjustments occurred half as frequently with use of Operative Armour in free flap reconstruction (1 adjustment per 10 stitches vs 1 adjustment per 5 stitches, P=.03) and a quarter as frequently in other breast reconstruction cases such as mastopexy (1 adjustment per 20 stitches vs 1 adjustment per 5 stitches, P=.002).
[CONCLUSIONS] Operative Armour effectively functions as an ESIP device by decreasing intraoperative needle passing and handling. Although sample size prohibits demonstrating a decrease in NSIs during observed cases, by decreasing behaviors that drive NSI risk, we anticipate an associated decrease in NSIs with use of the device.
[OBJECTIVE] We characterize Operative Armour's ESIP device effectiveness in a tertiary hospital, hypothesizing that this device will decrease NSI risk by decreasing behaviors associated with NSIs: needle passing and handling.
[METHODS] A prospective case-control study was conducted with institutional review board quality improvement designation in which authors observed skin closures of plastic surgery procedures. To ensure accuracy, one surgeon was observed at a time. Control surgeries were purely observational; intervention cases involved surgeon use of the device during skin closure. Outcomes of interest included needle passing, needle handling, lost needles, and loaded waiting needles.
[RESULTS] Surgeons were observed in 50 control and 50 intervention cases. Operative Armour eliminated needle passing during skin closure. One NSI occurred in one control case; no NSIs were observed in intervention cases (P=.36). The mean number of loaded and unprotected waiting needles was also significantly decreased in the intervention group from 2.3 to 0.2 (P<.001). Furthermore, a multivariable linear regression established that Operative Armour significantly decreased the number of needle adjustments by hand per stitch observed (F=3.72; P=.01). In fact, needle adjustments by hand decreased overall (1 adjustment per 10 stitches vs 1 adjustment per 5 stitches, P=.004), and adjustments occurred half as frequently with use of Operative Armour in free flap reconstruction (1 adjustment per 10 stitches vs 1 adjustment per 5 stitches, P=.03) and a quarter as frequently in other breast reconstruction cases such as mastopexy (1 adjustment per 20 stitches vs 1 adjustment per 5 stitches, P=.002).
[CONCLUSIONS] Operative Armour effectively functions as an ESIP device by decreasing intraoperative needle passing and handling. Although sample size prohibits demonstrating a decrease in NSIs during observed cases, by decreasing behaviors that drive NSI risk, we anticipate an associated decrease in NSIs with use of the device.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | mastopexy
|
유방성형술 | dict | 1 | |
| 시술 | free flap
|
피판재건술 | dict | 1 | |
| 해부 | breast
|
유방 | dict | 1 | |
| 해부 | skin
|
scispacy | 1 | ||
| 해부 | flap
|
scispacy | 1 | ||
| 합병증 | forearm
|
scispacy | 1 | ||
| 약물 | NSI
|
scispacy | 1 | ||
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | NSIs
→ needlestick injuries
|
C0085178
Needlestick Injuries
|
scispacy | 1 | |
| 질환 | NSI
|
scispacy | 1 | ||
| 질환 | decreasing behaviors
|
scispacy | 1 | ||
| 기타 | NSI
|
scispacy | 1 | ||
| 기타 | F=3.72
|
scispacy | 1 |
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.