Poly Implant Prothèse™ (PIP) experience in the United Kingdom: A prospective cohort study into the accuracy of diagnostic imaging findings in comparison to operative findings of 1029 implants.
Abstract
The Poly Implant Prosthèse™ (PIP) implants were withdrawn from market use in the United Kingdom on 31st March 2010 following Government issued advice. In June 2012 a final Government report was issued and during this period the majority of patients elected to have their prostheses removed. This study presents the operative findings of three surgeons. 517 patients were identified retrospectively from the implant database as having received PIP implants with a total of 1029 implants. 62 patients (124 implants) declined explantation after consultation and imaging. The data was recorded prospectively for all patients and included the clinical, imaging and operative findings. A total of 905 implants were removed of which 129 were ruptured at the time of explantation (14.25%). 27 implants were intact but the presence of liquid surrounding the prosthesis was noted. 93 implants were reported as being ruptured after diagnostic imaging but were intact operatively resulting in a test sensitivity of 0.82 and a specificity of 0.92 yielding a positive predictive value of 0.59 and a negative predictive value of 0.97 overall. Capsule formation was noted in 27 breasts (3%). Our study showed that the prevalence of PIP ruptures is comparable to other manufacturers. The prevalence of implant rupture predictably increased over time and the prevalence of abnormal capsule formation was similar to other manufacturers.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 합병증 | implant rupture
|
보형물 파열 | dict | 1 | |
| 약물 | Poly Implant
|
scispacy | 1 | ||
| 약물 | PIP
→ Poly Implant Prothèse™
|
scispacy | 1 | ||
| 약물 | Poly Implant Prosthèse
|
scispacy | 1 | ||
| 질환 | Capsule
|
scispacy | 1 | ||
| 기타 | patients
|
scispacy | 1 | ||
| 기타 | PIP
→ Poly Implant Prothèse™
|
scispacy | 1 |
MeSH Terms
Breast Implants; Diagnostic Imaging; Female; Humans; Prospective Studies; Prosthesis Design; Prosthesis Failure; Rupture; Sensitivity and Specificity; Silicone Gels; United Kingdom
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Five-year Results of the PERLE Implant in Breast Augmentation and Augmentation-mastopexy.
- High-Resolution Ultrasound and Postoperative Surveillance of the Augmented Breast.
- Surgical Treatment of Capsular Contracture with Acellular Dermal Matrix: 100 Consecutive Cases.
- Breast Implant Rupture Is Strongly Associated with Capsular Contracture: An Intrapatient Study.
- ACR Appropriateness Criteria® Breast Implant Evaluation: Update 2025.