Clinical accuracy and reproducibility of Portrait 3D Surgical Simulation Platform in breast augmentation.

Aesthetic surgery journal 2013 Vol.33(1) p. 84-92

Mailey B, Freel A, Wong R, Pointer DT, Khoobehi K

관련 도메인

Abstract

[BACKGROUND] Three-dimensional (3D) digital photography uses integrated image capture technology and rendering software to create 3D models. Volumetric measurements project simulated post-operative images prior to breast augmentation.

[OBJECTIVES] To evaluate the accuracy and reproducibility of breast volume measurements using the Portrait 3D Surgical Simulation Platform (Axis Three, Boston, Massachusetts).

[METHODS] Twenty-two patients underwent breast augmentation. 3D volumetric imaging analysis was performed by two independent observers preoperatively and at 6 weeks postoperatively. Simulated volumes were compared to actual implant volume using the Student's t test. Intra-observer reliability was evaluated by measuring internal consistency (Cronbach and 95% confidence interval [CI]) and test-retest reliability (intraclass correlation coefficient [ICC]) from the two observers' calculated volumes.

[RESULTS] Half (n = 11) of the patients received silicone implants and half saline; all were placed in the submuscular plane through an inframammary incision. No difference in volume estimation in preoperative or postoperative images (P = .49 and P = .14; and P = 1.0 and P = .37, in right and left breasts, respectively) was observed. The test-retest reliability between observers was excellent (ICC, 0.98; P < .001) and Cronbach's value (0.99; 95% CI 0.97-0.99; P < .001) demonstrated an excellent correlation. Regarding accuracy, difference in volume estimation between actual and simulated volumes varied between 0 to 106 mL (0 - 30%), with an absolute mean difference of 12.2% (42.5 mL).

[CONCLUSIONS] The Portrait 3D breast imaging system provides a highly reproducible 3D tool for measuring breast volume and simulating breast augmentation. Accuracy of the 3D models can vary up to 30% (mean 12.2%). This variability should be accounted for when using this technology to visually communicate with patients.

추출된 의학 개체 (NER)

유형영어 표현한국어 / 풀이UMLS CUI출처등장
해부 breast 유방 dict 7
시술 breast augmentation 유방성형술 dict 4
해부 inframammary scispacy 1
약물 silicone C0037114
silicones
scispacy 1
약물 [BACKGROUND] Three-dimensional scispacy 1
약물 [OBJECTIVES] scispacy 1
약물 saline scispacy 1
약물 [CONCLUSIONS] scispacy 1
기법 submuscular 근막하 평면 dict 1
질환 Axis C0004457
Axis vertebra
scispacy 1
기타 patients scispacy 1

MeSH Terms

Breast; Female; Humans; Imaging, Three-Dimensional; Mammaplasty; Photography; Reproducibility of Results

🔗 함께 등장하는 도메인

이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들

관련 논문