Form stability of the Style 410 anatomically shaped cohesive silicone gel-filled breast implant in subglandular breast augmentation evaluated with magnetic resonance imaging.
Abstract
[BACKGROUND] In this study, the authors evaluated the form stability of the Style 410 anatomically shaped cohesive silicone gel-filled breast implant after subglandular implantation.
[METHODS] Nine women who had undergone bilateral subglandular breast augmentation using Style 410 implants were examined in the prone and supine positions with magnetic resonance imaging.
[RESULTS] In the supine position, the dimensions of the implants were similar to those specified by the manufacturer. In the prone position, there was a mean increase of 29.5 percent (range, 9.7 to 53.3 percent) in implant projection. All implants had their point of maximal projection positioned in the lower pole in both positions.
[CONCLUSIONS] The Style 410 implant is described as a form-stable, anatomically shaped, cohesive, silicone gel-filled implant. However, its shell and the degree of silicone gel cohesivity allow for a change in form depending on body posture. Compared with the supine position, there was a marked increase in implant projection in the prone position. The Style 410 implant keeps its lower pole fullness after subglandular implantation in both the supine and prone positions.
[METHODS] Nine women who had undergone bilateral subglandular breast augmentation using Style 410 implants were examined in the prone and supine positions with magnetic resonance imaging.
[RESULTS] In the supine position, the dimensions of the implants were similar to those specified by the manufacturer. In the prone position, there was a mean increase of 29.5 percent (range, 9.7 to 53.3 percent) in implant projection. All implants had their point of maximal projection positioned in the lower pole in both positions.
[CONCLUSIONS] The Style 410 implant is described as a form-stable, anatomically shaped, cohesive, silicone gel-filled implant. However, its shell and the degree of silicone gel cohesivity allow for a change in form depending on body posture. Compared with the supine position, there was a marked increase in implant projection in the prone position. The Style 410 implant keeps its lower pole fullness after subglandular implantation in both the supine and prone positions.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 4 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 2 | |
| 해부 | subglandular
|
scispacy | 1 | ||
| 합병증 | subglandular breast
|
scispacy | 1 | ||
| 약물 | silicone
|
C0037114
silicones
|
scispacy | 1 | |
| 약물 | [BACKGROUND] In
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | silicone gel-filled breast implant
|
C0741713
Silicone gel-filled breast implant
|
scispacy | 1 | |
| 질환 | silicone gel-filled
|
scispacy | 1 | ||
| 질환 | fullness
|
C0439650
Fullness
|
scispacy | 1 | |
| 기타 | women
|
scispacy | 1 | ||
| 기타 | bilateral subglandular
|
scispacy | 1 | ||
| 기타 | form-stable
|
scispacy | 1 |
MeSH Terms
Adult; Breast; Breast Implants; Female; Humans; Magnetic Resonance Imaging; Mammaplasty; Middle Aged; Prone Position; Silicone Gels; Supine Position
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.