In vitro deflation of prefilled saline breast implants.
Abstract
[BACKGROUND] The purpose of this study was to determine whether or not prefilled breast implants retain their volume in packaging.
[METHODS] This study examined 32 Poly Implant Prosthesis prefilled textured saline breast implants. All of these implants were within the manufacturer's expiration date at the time of weighing. No holes were visible on any of the implants. All were weighed on an electronic scale. The measured weight was compared with the expected weight (based on the implant size as specified by the manufacturer) and the percentage deflation was calculated. The manufacturer declined to provide specific information about the manufacture dates of the implants; thus, relative age (rather than absolute age) was examined with respect to percentage deflation.
[RESULTS] Of the 32 implants examined, all showed some degree of deflation (range, 8.84 to 57.14 percent; 95 percent confidence interval of the mean, 22.01 +/- 4.17 percent). There was a moderate correlation (r = 0.41) between relative age of the implant and percentage deflation.
[CONCLUSIONS] From these results, it is clear that this type of prefilled saline breast implant does not maintain its volume in vitro. If these implants are used, the underfilling could contribute to a higher deflation rate and cosmetic deformity. This risk should be taken into account by plastic surgeons who use this type of implant in breast augmentation procedures.
[METHODS] This study examined 32 Poly Implant Prosthesis prefilled textured saline breast implants. All of these implants were within the manufacturer's expiration date at the time of weighing. No holes were visible on any of the implants. All were weighed on an electronic scale. The measured weight was compared with the expected weight (based on the implant size as specified by the manufacturer) and the percentage deflation was calculated. The manufacturer declined to provide specific information about the manufacture dates of the implants; thus, relative age (rather than absolute age) was examined with respect to percentage deflation.
[RESULTS] Of the 32 implants examined, all showed some degree of deflation (range, 8.84 to 57.14 percent; 95 percent confidence interval of the mean, 22.01 +/- 4.17 percent). There was a moderate correlation (r = 0.41) between relative age of the implant and percentage deflation.
[CONCLUSIONS] From these results, it is clear that this type of prefilled saline breast implant does not maintain its volume in vitro. If these implants are used, the underfilling could contribute to a higher deflation rate and cosmetic deformity. This risk should be taken into account by plastic surgeons who use this type of implant in breast augmentation procedures.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 5 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 1 | |
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 약물 | Poly Implant
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | breast implant
|
C0178391
breast implant procedure
|
scispacy | 1 | |
| 질환 | underfilling
|
scispacy | 1 | ||
| 질환 | cosmetic deformity
|
scispacy | 1 |
MeSH Terms
Breast Implants; Female; Humans; In Vitro Techniques; Prosthesis Failure; Sodium Chloride; Time Factors
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.