Mechanical and sensory testing of overfilled breast implants.
Abstract
[BACKGROUND] To obtain an ideal outcome in breast augmentation procedures, breast implants may be overfilled to reduce wrinkles and improve longevity. Determination of the optimal volume for saline-filled breast implants remains problematic for plastic surgeons.
[OBJECTIVE] We sought to evaluate the effect of overfilling or underfilling implants on sensory response and on implant shell mechanical properties.
[METHODS] Fifty-three Mentor Style 1600 smooth saline-filled mammary implants (275-cc base volume) were examined. Saline fill volumes ranged from 230 cc to 1650 cc. Implants were incubated at 37 degrees C and 100% humidity for 6 weeks. Before destructive testing to determine strength, elasticity, and toughness, 22 women ranging in age from 18 to 55 years graded the implants for firmness to palpation. Objective indices of firmness were measured by durometer.
[RESULTS] Implants filled 50% above manufacturer's recommended volume were considered firm, whereas implants 50% below manufacturer's recommended volume were considered natural or soft. A linear relationship was observed between sensory response and durometer hardness data. Fill volume did not affect shell mechanical properties: shell strength, toughness, and elasticity remained unchanged (P > .05).
[CONCLUSIONS] Saline-filled breast implants can be safely filled far beyond manufacturer's recommendations with regard to the mechanical properties of the Silastic polymer implant shell. Clinical judgment is required to avoid hardness due to overfill. Repeating this study using an in vivo incubation method is required to determine how overfill will affect implant shells over time in the harsh milieu of a biologic environment.
[OBJECTIVE] We sought to evaluate the effect of overfilling or underfilling implants on sensory response and on implant shell mechanical properties.
[METHODS] Fifty-three Mentor Style 1600 smooth saline-filled mammary implants (275-cc base volume) were examined. Saline fill volumes ranged from 230 cc to 1650 cc. Implants were incubated at 37 degrees C and 100% humidity for 6 weeks. Before destructive testing to determine strength, elasticity, and toughness, 22 women ranging in age from 18 to 55 years graded the implants for firmness to palpation. Objective indices of firmness were measured by durometer.
[RESULTS] Implants filled 50% above manufacturer's recommended volume were considered firm, whereas implants 50% below manufacturer's recommended volume were considered natural or soft. A linear relationship was observed between sensory response and durometer hardness data. Fill volume did not affect shell mechanical properties: shell strength, toughness, and elasticity remained unchanged (P > .05).
[CONCLUSIONS] Saline-filled breast implants can be safely filled far beyond manufacturer's recommendations with regard to the mechanical properties of the Silastic polymer implant shell. Clinical judgment is required to avoid hardness due to overfill. Repeating this study using an in vivo incubation method is required to determine how overfill will affect implant shells over time in the harsh milieu of a biologic environment.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 5 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 1 | |
| 해부 | mammary
|
유방 | dict | 1 | |
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 약물 | Saline
|
scispacy | 1 | ||
| 약물 | [RESULTS] Implants
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS] Saline-filled breast
|
scispacy | 1 | ||
| 질환 | underfilling
|
scispacy | 1 | ||
| 질환 | hardness
|
C0018599
Hardness
|
scispacy | 1 | |
| 질환 | wrinkles
|
scispacy | 1 | ||
| 기타 | women
|
scispacy | 1 | ||
| 기타 | Silastic polymer implant shell
|
scispacy | 1 |
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.