Vibration Exposure Safety Guidelines for Surgeons Using Power-Assisted Liposuction (PAL).
Abstract
[BACKGROUND] As power-assisted liposuction (PAL) gains in popularity, plastic surgeons operating these devices experience occupational exposure to hand-transmitted vibration, which can result in hand-arm vibration syndrome, a debilitating neurovasculopathy.
[OBJECTIVES] The objective of the study was to determine vibration exposure from the utilization of a PAL device during surgery to generate recommendations for safe use.
[METHODS] Vibration emission of a commonly utilized PAL system (MicroAire-650, Surgical Instruments, Charlottesville, VA) was examined employing a vibration data logger under both controlled laboratory conditions and during 13 typical liposuction cases. Data were analyzed and compared with established safety limits of vibration exposure.
[RESULTS] The experiments demonstrated a mean vibration magnitude of typical liposuction surgeries to be 5.69 ± 0.77 m/s2 (range, 4.59-6.27 m/s2), which is significantly higher than the manufacturer declared value of 3.77 m/s2. Cannula size was shown to be the most significant contributor to vibration magnitude, with larger cannulas causing more vibration transmission.
[CONCLUSIONS] These results indicate that recommendations must be made to prevent undue occupational exposure to vibration from PAL. The MicroAire-650 can generally be safely utilized for less than 1.5 h/d. At exposure levels >1.5 h/d, there is increased risk of developing vibration-related injuries, and vibration-reducing strategies should be implemented. At exposure levels >6 h/d, the safety limit is exceeded and there is significantly increased risk of developing hand-arm vibration syndrome and vibration exposure should be halted.
[OBJECTIVES] The objective of the study was to determine vibration exposure from the utilization of a PAL device during surgery to generate recommendations for safe use.
[METHODS] Vibration emission of a commonly utilized PAL system (MicroAire-650, Surgical Instruments, Charlottesville, VA) was examined employing a vibration data logger under both controlled laboratory conditions and during 13 typical liposuction cases. Data were analyzed and compared with established safety limits of vibration exposure.
[RESULTS] The experiments demonstrated a mean vibration magnitude of typical liposuction surgeries to be 5.69 ± 0.77 m/s2 (range, 4.59-6.27 m/s2), which is significantly higher than the manufacturer declared value of 3.77 m/s2. Cannula size was shown to be the most significant contributor to vibration magnitude, with larger cannulas causing more vibration transmission.
[CONCLUSIONS] These results indicate that recommendations must be made to prevent undue occupational exposure to vibration from PAL. The MicroAire-650 can generally be safely utilized for less than 1.5 h/d. At exposure levels >1.5 h/d, there is increased risk of developing vibration-related injuries, and vibration-reducing strategies should be implemented. At exposure levels >6 h/d, the safety limit is exceeded and there is significantly increased risk of developing hand-arm vibration syndrome and vibration exposure should be halted.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | liposuction
|
지방흡입 | dict | 4 | |
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 약물 | [OBJECTIVES]
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | hand-arm vibration syndrome
|
C1363854
Hand-Arm Vibration Syndrome
|
scispacy | 1 | |
| 질환 | neurovasculopathy
|
scispacy | 1 | ||
| 질환 | vibration-related injuries
|
scispacy | 1 | ||
| 질환 | MicroAire-650
|
scispacy | 1 | ||
| 기타 | Charlottesville
|
scispacy | 1 |
MeSH Terms
Hand; Hand-Arm Vibration Syndrome; Humans; Lipectomy; Surgeons; Vibration
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Quantitative Assessment of Cannula Kinematics in Liposuction Surgical Procedures Using a Marker-Based Tracking System.
- Fat harvesting protocol for enhanced stem cell viability - pilot study.
- Scarless Infragluteal Fixation (SIF): Correction of Post-Liposuction Infragluteal Deformity.
- [Super microsurgical lymphaticovenular anastomosis for limb lymphedema: An outcome analysis based on clinical stage and indocyanine green pattern].
- Iatrogenic pneumothorax associated with surgeries at anatomically thoracic-adjacent and non-adjacent sites: case report and scoping review.