Feasibility of augmented reality combine patient-specific implants (PSI) applied to navigation in mandibular genioplasty: A phantom experiment.
Abstract
[PURPOSE] Genitoplasty is becoming more and more common, and it is important to improve the accuracy of the procedure and simplify the procedure. This experiment explores the feasibility of using augmented reality (AR) technology combined with PSI titanium plates for navigational assistance in genioplasty performed on models, aiming to study the precision of such surgical interventions.
[METHODS] Twelve genioplasty procedures were designed and implemented on 3D-printed resin mandibular models by the same surgeon using three different approaches: AR+3DT group (AR+PSI) , 3DT group (patient-specific titanium plate) , and a traditional free-hand group(FH group). Postoperative models were assessed using CBCT to evaluate surgical accuracy.
[RESULTS] In terms of osteotomy accuracy, the AR group demonstrated a surgical error of 0.9440±0.5441 mm, significantly lower than the control group, which had an error of 1.685±0.8907 mm (P < 0.0001). In experiments positioning the distal segment of the chin, the overall centroid shift in the AR group was 0.3661±0.1360 mm, significantly less than the 2.304±0.9629 mm in the 3DT group and 1.562±0.9799 mm in the FH group (P < 0.0001). Regarding angular error, the AR+3DT group showed 2.825±1.373°, significantly <8.283±3.640° in the 3DT group and 7.234±5.241° in the FH group.
[CONCLUSION] AR navigation technology combined with PSI titanium plates demonstrates higher surgical accuracy compared to traditional methods and shows feasibility for use. Further validation through clinical trials is necessary.
[METHODS] Twelve genioplasty procedures were designed and implemented on 3D-printed resin mandibular models by the same surgeon using three different approaches: AR+3DT group (AR+PSI) , 3DT group (patient-specific titanium plate) , and a traditional free-hand group(FH group). Postoperative models were assessed using CBCT to evaluate surgical accuracy.
[RESULTS] In terms of osteotomy accuracy, the AR group demonstrated a surgical error of 0.9440±0.5441 mm, significantly lower than the control group, which had an error of 1.685±0.8907 mm (P < 0.0001). In experiments positioning the distal segment of the chin, the overall centroid shift in the AR group was 0.3661±0.1360 mm, significantly less than the 2.304±0.9629 mm in the 3DT group and 1.562±0.9799 mm in the FH group (P < 0.0001). Regarding angular error, the AR+3DT group showed 2.825±1.373°, significantly <8.283±3.640° in the 3DT group and 7.234±5.241° in the FH group.
[CONCLUSION] AR navigation technology combined with PSI titanium plates demonstrates higher surgical accuracy compared to traditional methods and shows feasibility for use. Further validation through clinical trials is necessary.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | genioplasty
|
턱끝성형술 | dict | 3 | |
| 해부 | 3DT
|
scispacy | 1 | ||
| 해부 | group(FH
|
scispacy | 1 | ||
| 합병증 | mandibular genioplasty
|
scispacy | 1 | ||
| 약물 | group(FH
|
scispacy | 1 | ||
| 약물 | 7.234±5.241
|
scispacy | 1 | ||
| 질환 | Genitoplasty
|
scispacy | 1 | ||
| 질환 | AR+3DT
|
scispacy | 1 | ||
| 기타 | PSI
→ patient-specific implants
|
scispacy | 1 |
MeSH Terms
Humans; Genioplasty; Feasibility Studies; Augmented Reality; Mandible; Surgery, Computer-Assisted; Printing, Three-Dimensional; Titanium; Bone Plates; Cone-Beam Computed Tomography; Phantoms, Imaging; Male; Models, Anatomic
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Long-term effects of inferior alveolar nerve injury on the quality of life of orthognathic patients: a clinical assessment.
- Mapping the Mandibular Lingual Foramina for Safer Chin Surgery: CT Morphometry and Predictive Modeling.
- Effectiveness and Safety of a Polyvinyl Alcohol Microsphere and Hyaluronic Acid Suspension for Chin Augmentation: A Randomized Controlled Trial and Multidisciplinary Study.
- Safety and Effectiveness of Two High-G' Soft Tissue Fillers for Chin Augmentation: A Prospective, Randomized, Comparator-Controlled, Evaluator-Blinded Trial.
- Minimally Invasive Protocol for the Management of Unilateral Condylar Hyperplasia: Case Series on Seven Patients.