Effect of Field of View and Voxel Size on CBCT-Based Accuracy of Dynamic Navigation in Endodontic Microsurgery: An In Vitro Study.
Abstract
[INTRODUCTION] This study aimed to evaluate the influence of field of view (FOV) and voxel size on the accuracy of dynamic navigation (DN)-assisted endodontic microsurgery (EMS).
[METHODS] Nine sets of maxillary and mandibular 3-dimensional-printed jaw models composed of 180 teeth were divided into 9 groups with different FOVs (80 × 80 mm, 60 × 60 mm, and 40 × 40 mm) and voxel sizes (0.3 mm, 0.16 mm, and 0.08 mm). The endodontic DN system was used to plan and execute the EMS. The accuracy of the DN-EMS was represented by the platform deviation, end deviation, angular deviation, resection angle, and resection length deviation. Statistical analyses were performed using SPSS 24.0, and the significance level was set at P < .05.
[RESULTS] The average platform deviation, end deviation, angular deviation, resection angle, and resection length deviation were 0.69 ± 0.31 mm, 0.93 ± 0.44 mm, 3.47 ± 1.80°, 2.35 ± 1.76°, and 0.41 ± 0.29 mm, respectively. No statistically significant differences in accuracy were observed between the nine FOV and voxel size groups.
[CONCLUSIONS] FOV and voxel size did not appear to play an important role in the accuracy of DN-EMS. Considering the image quality and radiation dose, it is reasonable to select a limited FOV (such as 40 × 40 mm and 60 × 60 mm) to cover only the registration device, involved teeth, and periapical lesion. The voxel size should be selected according to the required resolution and cone-beam computed tomography units.
[METHODS] Nine sets of maxillary and mandibular 3-dimensional-printed jaw models composed of 180 teeth were divided into 9 groups with different FOVs (80 × 80 mm, 60 × 60 mm, and 40 × 40 mm) and voxel sizes (0.3 mm, 0.16 mm, and 0.08 mm). The endodontic DN system was used to plan and execute the EMS. The accuracy of the DN-EMS was represented by the platform deviation, end deviation, angular deviation, resection angle, and resection length deviation. Statistical analyses were performed using SPSS 24.0, and the significance level was set at P < .05.
[RESULTS] The average platform deviation, end deviation, angular deviation, resection angle, and resection length deviation were 0.69 ± 0.31 mm, 0.93 ± 0.44 mm, 3.47 ± 1.80°, 2.35 ± 1.76°, and 0.41 ± 0.29 mm, respectively. No statistically significant differences in accuracy were observed between the nine FOV and voxel size groups.
[CONCLUSIONS] FOV and voxel size did not appear to play an important role in the accuracy of DN-EMS. Considering the image quality and radiation dose, it is reasonable to select a limited FOV (such as 40 × 40 mm and 60 × 60 mm) to cover only the registration device, involved teeth, and periapical lesion. The voxel size should be selected according to the required resolution and cone-beam computed tomography units.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | microsurgery
|
미세수술 | dict | 2 | |
| 해부 | maxillary
|
scispacy | 1 | ||
| 해부 | jaw
|
scispacy | 1 | ||
| 해부 | teeth
|
scispacy | 1 | ||
| 해부 | FOVs
|
scispacy | 1 | ||
| 해부 | FOV
→ field of view
|
scispacy | 1 | ||
| 약물 | ± 0.31
|
C4517447
0.31
|
scispacy | 1 | |
| 약물 | [INTRODUCTION]
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS] FOV
|
scispacy | 1 | ||
| 기타 | mandibular
|
scispacy | 1 |
MeSH Terms
Microsurgery; Spiral Cone-Beam Computed Tomography; Research Design; Cone-Beam Computed Tomography; Maxilla
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.