Determination of residual acrylamide in medical polyacrylamide hydrogel by high performance liquid chromatography tandem mass spectroscopy.

Biomedical and environmental sciences : BES 2009 Vol.22(1) p. 28-31

Li WW, Li H, Liu ZF, Qiao Q

Abstract

[OBJECTIVE] To determine residual acrylamide in medical polyacrylamide hydrogel by high performance liquid chromatography tandem mass spectroscopy (HPLC-MS).

[METHODS] After (13)C3 labeled acrylamide was added, the sample was extracted with water and then cleaned up with Extrelut 20. The polyacrylamide hydrogel sample and 20 clinical cases were analyzed by HPLC-MS/MS and isotope dilution quantifying technique in selected reaction monitoring (SRM) mode.

[RESULTS] Acrylamide was separated from polyacrylamide hydrogel. The concentration of acrylamide in polyacrylamide hydrogel ranged from 3.9 x 10(-9) to 3.1 x 10(-8) g/L in the 20 clinical cases. The peak area was favorable linear and the range was up to 3000 microg/L. The recovery rate was 103.1% with a relative standard deviation (RSD) of 6.20%, when the mark level was 50 microg/L.

[CONCLUSION] HPLC-MS is a rapid, accurate, and sensitive method for the determination of residual acrylamide in medical polyacrylamide hydrogel.

추출된 의학 개체 (NER)

유형영어 표현한국어 / 풀이UMLS CUI출처등장
약물 acrylamide C0050587
acrylamide
scispacy 1
약물 [OBJECTIVE] scispacy 1
약물 [CONCLUSION] scispacy 1
질환 microg/L. scispacy 1

MeSH Terms

Acrylamide; Acrylic Resins; Biocompatible Materials; Calibration; Chromatography, High Pressure Liquid; Female; Gels; Humans; Materials Testing; Surgery, Plastic; Tandem Mass Spectrometry