A Wireless Near-Infrared Spectroscopy Device for Flap Monitoring: Proof of Concept in a Porcine Musculocutaneous Flap Model.

Journal of reconstructive microsurgery 2022 Vol.38(2) p. 96-105

Wu C, Rwei AY, Lee JY, Ouyang W, Jacobson L, Shen H, Luan H, Xu Y, Park JB, Kwak SS, Ni X, Bai W, Franklin D, Li S, Liu Y, Ni X, Westman AM, MacEwan MR, Rogers JA, Pet MA

관련 도메인

Abstract

[BACKGROUND]  Current near-infrared spectroscopy (NIRS)-based systems for continuous flap monitoring are highly sensitive for detecting malperfusion. However, the clinical utility and user experience are limited by the wired connection between the sensor and bedside console. This wire leads to instability of the flap-sensor interface and may cause false alarms.

[METHODS]  We present a novel wearable wireless NIRS sensor for continuous fasciocutaneous free flap monitoring. This waterproof silicone-encapsulated Bluetooth-enabled device contains two light-emitting diodes and two photodetectors in addition to a battery sufficient for 5 days of uninterrupted function. This novel device was compared with a ViOptix T.Ox monitor in a porcine rectus abdominus myocutaneous flap model of arterial and venous occlusions.

[RESULTS]  Devices were tested in four flaps using three animals. Both devices produced very similar tissue oxygen saturation (StO) tracings throughout the vascular clamping events, with obvious and parallel changes occurring on arterial clamping, arterial release, venous clamping, and venous release. Small interdevice variations in absolute StO value readings and magnitude of change were observed. The normalized cross-correlation at zero lag describing correspondence between the novel NIRS and T.Ox devices was >0.99 in each trial.

[CONCLUSION]  The wireless NIRS flap monitor is capable of detecting StO changes resultant from arterial vascular occlusive events. In this porcine flap model, the functionality of this novel sensor closely mirrored that of the T.Ox wired platform. This device is waterproof, highly adhesive, skin conforming, and has sufficient battery life to function for 5 days. Clinical testing is necessary to determine if this wireless functionality translates into fewer false-positive alarms and a better user experience.

추출된 의학 개체 (NER)

유형영어 표현한국어 / 풀이UMLS CUI출처등장
시술 flap 피판재건술 dict 7
시술 free flap 피판재건술 dict 1
해부 StO → saturation scispacy 1
해부 skin scispacy 1
합병증 fasciocutaneous free scispacy 1
합병증 flaps scispacy 1
약물 oxygen C0030054
oxygen
scispacy 1
약물 [BACKGROUND] scispacy 1
질환 malperfusion scispacy 1
질환 arterial vascular occlusive scispacy 1
질환 StO → saturation scispacy 1
기타 Porcine Musculocutaneous Flap scispacy 1
기타 porcine rectus abdominus myocutaneous flap scispacy 1
기타 arterial scispacy 1
기타 venous scispacy 1
기타 tissue oxygen scispacy 1
기타 vascular scispacy 1
기타 arterial vascular scispacy 1
기타 porcine flap scispacy 1

MeSH Terms

Animals; Free Tissue Flaps; Monitoring, Physiologic; Myocutaneous Flap; Oxygen; Spectroscopy, Near-Infrared; Swine; Veins

🔗 함께 등장하는 도메인

이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들

관련 논문