Data analytics interrogates robotic surgical performance using a microsurgery-specific haptic device.
Abstract
[OBJECTIVES] With the increase in robot-assisted cases, recording the quantifiable dexterity of surgeons is essential for proficiency evaluations. The present study employs sensor-based kinematics and recorded surgeon experience for evaluating a new haptic device.
[METHODS] Thirty surgeons performed a task simulating micromanipulation with neuroArmPLUS and two commercially available hand-controllers. The surgical performance was evaluated based on subjective measures obtained from survey and objective features derived from the sensors. Statistical analyses were performed to assess the hand-controllers and regression analysis was used to identify the key features and develop a machine learning model for surgical skill assessment.
[FINDINGS] MANCOVA tests on objective features demonstrated significance (α = 0.05) for time (p = 0.02), errors (p = 0.01), distance (p = 0.03), clutch incidents (p = 0.03), and forces (p = 0.00). The majority of metrics were in favor of neuroArmPLUS. The surgeons found it smoother, more comfortable, less tiring, and easier to maneuver with more realistic force feedback. The ensemble machine learning model trained with 5-fold cross-validation showed an accuracy (SD) of 0.78 (0.15) in surgeon skill classification.
[CONCLUSIONS] This study validates the importance of incorporating a superior haptic device in telerobotic surgery for standardization of surgical education and patient care.
[METHODS] Thirty surgeons performed a task simulating micromanipulation with neuroArmPLUS and two commercially available hand-controllers. The surgical performance was evaluated based on subjective measures obtained from survey and objective features derived from the sensors. Statistical analyses were performed to assess the hand-controllers and regression analysis was used to identify the key features and develop a machine learning model for surgical skill assessment.
[FINDINGS] MANCOVA tests on objective features demonstrated significance (α = 0.05) for time (p = 0.02), errors (p = 0.01), distance (p = 0.03), clutch incidents (p = 0.03), and forces (p = 0.00). The majority of metrics were in favor of neuroArmPLUS. The surgeons found it smoother, more comfortable, less tiring, and easier to maneuver with more realistic force feedback. The ensemble machine learning model trained with 5-fold cross-validation showed an accuracy (SD) of 0.78 (0.15) in surgeon skill classification.
[CONCLUSIONS] This study validates the importance of incorporating a superior haptic device in telerobotic surgery for standardization of surgical education and patient care.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | microsurgery
|
미세수술 | dict | 1 | |
| 해부 | clutch
|
scispacy | 1 | ||
| 약물 | [OBJECTIVES]
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 기법 | robot-assisted
|
로봇수술 | dict | 1 | |
| 기타 | patient
|
scispacy | 1 |
MeSH Terms
Adult; Data Science; Humans; Microsurgery; Middle Aged; Neurosurgery; Robotic Surgical Procedures; Surgeons; Time Factors; Touch; Young Adult
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.