Real-Time Monitoring of Blood Pressure Using Digitalized Pulse Arrival Time Calculation Technology for Prompt Detection of Sudden Hypertensive Episodes During Laryngeal Microsurgery: Retrospective Observational Study.
Abstract
[BACKGROUND] Laryngeal microsurgery (LMS) is often accompanied by a sudden increase in blood pressure (BP) during surgery because of stimulation around the larynx. This sudden change in the hemodynamic status is not immediately reflected in a casual cuff-type measurement that takes intermittent readings every 3 to 5 min.
[OBJECTIVE] This study aimed to investigate the potential of pulse arrival time (PAT) as a marker for a BP surge, which usually occurs in patients undergoing LMS.
[METHODS] Intermittent measurements of BP and electrocardiogram (ECG) and photoplethysmogram (PPG) signals were recorded during LMS. PAT was defined as the interval between the R-peak on the ECG and the maximum slope on the PPG. Mean PAT values before and after BP increase were compared. PPG-related parameters and the correlations between changes in these variables were calculated.
[RESULTS] BP surged because of laryngoscopic manipulation (mean systolic BP [SBP] from 115.3, SD 21.4 mmHg, to 159.9, SD 25.2 mmHg; P<.001), whereas PAT decreased significantly (from mean 460.6, SD 51.9 ms, to 405.8, SD 50.1 ms; P<.001) in most of the cases. The change in SBP showed a significant correlation with the inverse of the PAT (r=0.582; P<.001). Receiver-operating characteristic curve analysis indicated that an increase of 11.5% in the inverse of the PAT could detect a 40% increase in SBP, and the area under the curve was 0.814.
[CONCLUSIONS] During LMS, where invasive arterial catheterization is not always possible, PAT shows good correlation with SBP and may, therefore, have the potential to identify abrupt BP surges during laryngoscopic manipulations in a noninvasive manner.
[OBJECTIVE] This study aimed to investigate the potential of pulse arrival time (PAT) as a marker for a BP surge, which usually occurs in patients undergoing LMS.
[METHODS] Intermittent measurements of BP and electrocardiogram (ECG) and photoplethysmogram (PPG) signals were recorded during LMS. PAT was defined as the interval between the R-peak on the ECG and the maximum slope on the PPG. Mean PAT values before and after BP increase were compared. PPG-related parameters and the correlations between changes in these variables were calculated.
[RESULTS] BP surged because of laryngoscopic manipulation (mean systolic BP [SBP] from 115.3, SD 21.4 mmHg, to 159.9, SD 25.2 mmHg; P<.001), whereas PAT decreased significantly (from mean 460.6, SD 51.9 ms, to 405.8, SD 50.1 ms; P<.001) in most of the cases. The change in SBP showed a significant correlation with the inverse of the PAT (r=0.582; P<.001). Receiver-operating characteristic curve analysis indicated that an increase of 11.5% in the inverse of the PAT could detect a 40% increase in SBP, and the area under the curve was 0.814.
[CONCLUSIONS] During LMS, where invasive arterial catheterization is not always possible, PAT shows good correlation with SBP and may, therefore, have the potential to identify abrupt BP surges during laryngoscopic manipulations in a noninvasive manner.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | microsurgery
|
미세수술 | dict | 2 | |
| 해부 | Blood
|
scispacy | 1 | ||
| 해부 | larynx
|
scispacy | 1 | ||
| 해부 | cuff-type
|
scispacy | 1 | ||
| 해부 | SBP
|
scispacy | 1 | ||
| 합병증 | Laryngeal Microsurgery
|
scispacy | 1 | ||
| 약물 | PAT
→ pulse arrival time
|
scispacy | 1 | ||
| 약물 | [BACKGROUND] Laryngeal microsurgery
|
scispacy | 1 | ||
| 약물 | [OBJECTIVE]
|
scispacy | 1 | ||
| 약물 | photoplethysmogram
|
scispacy | 1 | ||
| 약물 | [RESULTS] BP
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | Hypertensive
|
C0857121
Hypertensive (finding)
|
scispacy | 1 | |
| 질환 | LMS
→ Laryngeal microsurgery
|
scispacy | 1 | ||
| 질환 | sudden increase
|
scispacy | 1 | ||
| 질환 | ECG
→ electrocardiogram
|
scispacy | 1 | ||
| 기타 | patients
|
scispacy | 1 | ||
| 기타 | SD 25.2
|
scispacy | 1 | ||
| 기타 | PAT
→ pulse arrival time
|
scispacy | 1 | ||
| 기타 | arterial
|
scispacy | 1 |
MeSH Terms
Blood Pressure; Blood Pressure Determination; Female; Heart Rate; Humans; Hypertension; Larynx; Male; Microsurgery; Middle Aged; Retrospective Studies
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.