A Novel Technology for Free Flap Monitoring: Pilot Study of a Wireless, Biodegradable Sensor.
Abstract
[BACKGROUND] Accurate monitoring of free flap perfusion after complex reconstruction is critical for early recognition of flap compromise. Surgeons use a variety of subjective and objective measures to evaluate flap perfusion postoperatively. However, these measures have some limitations. We have developed a wireless, biodegradable, and flexible sensor that can be applied to real-time postoperative free flap monitoring. Here we assess the biocompatibility and function of our novel sensor.
[METHODS] Seven Sprague-Dawley (SD) rats were used for biocompatibility studies. The sensor was implanted around the femoral artery near the inguinal ligament on one leg (implant side) and sham surgery was performed on the contralateral leg (control side). At 6 and 12 weeks, samples were harvested to assess the inflammation within and around the implant and artery. Two animals were used to assess sensor function. Sensor function was evaluated at implantation and 7 days after the implantation. Signal changes after venous occlusion were also assessed in an epigastric artery island flap model.
[RESULTS] In biocompatibility studies, the diameter of the arterial lumen and intima thickness in the implant group were not significantly different than the control group at the 12-week time point. The number of CD-68 positive cells that infiltrated into the soft tissue, surrounding the femoral artery, was also not significantly different between groups at the 12-week time point. For sensor function, accurate signaling could be recorded at implantation and 7 days later. A change in arterial signal was noted immediately after venous occlusion in a flap model.
[CONCLUSION] The novel wireless, biodegradable sensor presented here is biocompatible and capable of detecting arterial blood flow and venous occlusion with high sensitivity. This promising new technology could combat the complications of wired sensors, while improving the survival rate of flaps with vessel compromise due to its responsive nature.
[METHODS] Seven Sprague-Dawley (SD) rats were used for biocompatibility studies. The sensor was implanted around the femoral artery near the inguinal ligament on one leg (implant side) and sham surgery was performed on the contralateral leg (control side). At 6 and 12 weeks, samples were harvested to assess the inflammation within and around the implant and artery. Two animals were used to assess sensor function. Sensor function was evaluated at implantation and 7 days after the implantation. Signal changes after venous occlusion were also assessed in an epigastric artery island flap model.
[RESULTS] In biocompatibility studies, the diameter of the arterial lumen and intima thickness in the implant group were not significantly different than the control group at the 12-week time point. The number of CD-68 positive cells that infiltrated into the soft tissue, surrounding the femoral artery, was also not significantly different between groups at the 12-week time point. For sensor function, accurate signaling could be recorded at implantation and 7 days later. A change in arterial signal was noted immediately after venous occlusion in a flap model.
[CONCLUSION] The novel wireless, biodegradable sensor presented here is biocompatible and capable of detecting arterial blood flow and venous occlusion with high sensitivity. This promising new technology could combat the complications of wired sensors, while improving the survival rate of flaps with vessel compromise due to its responsive nature.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | flap
|
피판재건술 | dict | 4 | |
| 시술 | free flap
|
피판재건술 | dict | 3 | |
| 해부 | leg
|
scispacy | 1 | ||
| 해부 | arterial lumen
|
scispacy | 1 | ||
| 해부 | intima
|
scispacy | 1 | ||
| 해부 | cells
|
scispacy | 1 | ||
| 해부 | soft tissue
|
scispacy | 1 | ||
| 합병증 | leg (
|
scispacy | 1 | ||
| 합병증 | arterial blood
|
scispacy | 1 | ||
| 합병증 | flaps
|
scispacy | 1 | ||
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 질환 | inflammation
|
C0021368
Inflammation
|
scispacy | 1 | |
| 질환 | venous occlusion
|
C0948441
Venoocclusive disease
|
scispacy | 1 | |
| 기타 | Sprague-Dawley (SD) rats
|
scispacy | 1 | ||
| 기타 | femoral artery
|
scispacy | 1 | ||
| 기타 | inguinal ligament
|
scispacy | 1 | ||
| 기타 | artery
|
scispacy | 1 | ||
| 기타 | venous
|
scispacy | 1 | ||
| 기타 | epigastric artery
|
scispacy | 1 | ||
| 기타 | CD-68
|
scispacy | 1 | ||
| 기타 | arterial
|
scispacy | 1 | ||
| 기타 | vessel
|
scispacy | 1 |
MeSH Terms
Animals; Biocompatible Materials; Free Tissue Flaps; Hindlimb; Microsurgery; Monitoring, Physiologic; Pilot Projects; Rats; Rats, Sprague-Dawley; Regional Blood Flow; Wireless Technology
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.