Baker Grade IV Capsular Contracture Is Correlated with an Increased Amount of Silicone Material: An Intrapatient Study.
Abstract
[BACKGROUND] Breast implant surgery is one of the most frequently performed procedures by plastic surgeons worldwide. However, the relationship between silicone leakage and the most common complication, capsular contracture, is far from understood. This study aimed to compare Baker grade I with Baker grade IV capsules regarding their silicone content in an intradonor setting, using two previously validated imaging techniques.
[METHODS] Twenty-two donor-matched capsules from 11 patients experiencing unilateral complaints were included after bilateral explantation surgery. All capsules were examined using both stimulated Raman scattering (SRS) imaging and staining with modified oil red O (MORO). Evaluation was done visually for qualitative and semiquantitative assessment and automated for quantitative analysis.
[RESULTS] Using both SRS and MORO techniques, silicone was found in more Baker grade IV capsules (eight of 11 and 11 of 11, respectively) than in Baker grade I capsules (three of 11 and five of 11, respectively). Baker grade IV capsules also showed significantly more silicone content compared with the Baker grade I capsules. This was true for semiquantitative assessment for both SRS and MORO techniques ( P = 0.019 and P = 0.006, respectively), whereas quantitative analysis proved to be significant for MORO alone ( P = 0.026 versus P = 0.248 for SRS, respectively).
[CONCLUSIONS] In this study, a significant correlation between capsule silicone content and capsular contracture is shown. An extensive and continued foreign body response to silicone particles is likely to be responsible. Considering the widespread use of silicone breast implants, these results affect many women worldwide and warrant a more focused research effort.
[CLINICAL QUESTION/LEVEL OF EVIDENCE] Risk, III.
[METHODS] Twenty-two donor-matched capsules from 11 patients experiencing unilateral complaints were included after bilateral explantation surgery. All capsules were examined using both stimulated Raman scattering (SRS) imaging and staining with modified oil red O (MORO). Evaluation was done visually for qualitative and semiquantitative assessment and automated for quantitative analysis.
[RESULTS] Using both SRS and MORO techniques, silicone was found in more Baker grade IV capsules (eight of 11 and 11 of 11, respectively) than in Baker grade I capsules (three of 11 and five of 11, respectively). Baker grade IV capsules also showed significantly more silicone content compared with the Baker grade I capsules. This was true for semiquantitative assessment for both SRS and MORO techniques ( P = 0.019 and P = 0.006, respectively), whereas quantitative analysis proved to be significant for MORO alone ( P = 0.026 versus P = 0.248 for SRS, respectively).
[CONCLUSIONS] In this study, a significant correlation between capsule silicone content and capsular contracture is shown. An extensive and continued foreign body response to silicone particles is likely to be responsible. Considering the widespread use of silicone breast implants, these results affect many women worldwide and warrant a more focused research effort.
[CLINICAL QUESTION/LEVEL OF EVIDENCE] Risk, III.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 합병증 | capsular contracture
|
피막구축 | dict | 3 | |
| 해부 | breast
|
유방 | dict | 2 | |
| 해부 | oil red O
|
scispacy | 1 | ||
| 약물 | Silicone
|
C0037114
silicones
|
scispacy | 1 | |
| 약물 | silicone leakage
|
scispacy | 1 | ||
| 약물 | [BACKGROUND] Breast implant
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | Contracture
|
C0009917
Contracture
|
scispacy | 1 | |
| 질환 | SRS
→ stimulated Raman scattering
|
scispacy | 1 | ||
| 질환 | silicone breast implants
|
C0741713
Silicone gel-filled breast implant
|
scispacy | 1 | |
| 질환 | Baker grade I capsules
|
scispacy | 1 | ||
| 질환 | capsule
|
scispacy | 1 | ||
| 기타 | Material
|
scispacy | 1 | ||
| 기타 | capsular
|
scispacy | 1 | ||
| 기타 | patients
|
scispacy | 1 | ||
| 기타 | Baker grade
|
scispacy | 1 | ||
| 기타 | women
|
scispacy | 1 |
MeSH Terms
Humans; Female; Silicones; Breast Implants; Breast Implantation; Device Removal; Contracture; Implant Capsular Contracture; Silicone Gels
📑 인용 관계
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- The Plastic Surgery In-Service Examination: A Scoping Review.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.