Ten-year Core Study Data for Sientra's Food and Drug Administration-Approved Round and Shaped Breast Implants with Cohesive Silicone Gel.
Abstract
[BACKGROUND] In March 2012, the United States Food and Drug Administration approved Sientra's application for premarket approval of its portfolio of round and shaped silicone gel breast implants based on data from the largest silicone gel breast implant study to date. This article presents the results of Sientra's Core Study at the conclusion of 10 years.
[METHODS] The Sientra Core Study was a 10-year, open-label, prospective, multicenter clinical study designed to assess the safety and effectiveness of Sientra's breast implants in augmentation and reconstruction. A total of 1,788 patients (3,506 implants) were enrolled, including 1,116 primary augmentation, 363 revision-augmentation, 225 primary reconstruction, and 84 revision-reconstruction.
[RESULTS] Across all cohorts, the rate of rupture by patient was 8.6%, the rate of Baker grade III/IV capsular contracture was 13.5%, and the rate of reoperation was 31.5%. The rate of capsular contracture was statistically significantly lower for textured devices [9.0%; 95% confidence interval (CI), 7.0-11.5%] compared with smooth devices (17.5%; 95% CI, 14.9-20.4%). There were no cases of breast implant-associated anaplastic large cell lymphoma. Primary reasons for reoperations included capsular contracture (18.8%), and style/size change (19.3%), with over 50% of the reoperations due to cosmetic reasons.
[CONCLUSION] The 10-year results of Sientra's Core Study support a comprehensive safety and effectiveness profile of Sientra's portfolio of round and shaped breast implants.
[METHODS] The Sientra Core Study was a 10-year, open-label, prospective, multicenter clinical study designed to assess the safety and effectiveness of Sientra's breast implants in augmentation and reconstruction. A total of 1,788 patients (3,506 implants) were enrolled, including 1,116 primary augmentation, 363 revision-augmentation, 225 primary reconstruction, and 84 revision-reconstruction.
[RESULTS] Across all cohorts, the rate of rupture by patient was 8.6%, the rate of Baker grade III/IV capsular contracture was 13.5%, and the rate of reoperation was 31.5%. The rate of capsular contracture was statistically significantly lower for textured devices [9.0%; 95% confidence interval (CI), 7.0-11.5%] compared with smooth devices (17.5%; 95% CI, 14.9-20.4%). There were no cases of breast implant-associated anaplastic large cell lymphoma. Primary reasons for reoperations included capsular contracture (18.8%), and style/size change (19.3%), with over 50% of the reoperations due to cosmetic reasons.
[CONCLUSION] The 10-year results of Sientra's Core Study support a comprehensive safety and effectiveness profile of Sientra's portfolio of round and shaped breast implants.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 6 | |
| 합병증 | capsular contracture
|
피막구축 | dict | 3 | |
| 해부 | smooth
|
scispacy | 1 | ||
| 합병증 | breast implants
|
scispacy | 1 | ||
| 합병증 | breast implant
|
scispacy | 1 | ||
| 합병증 | anaplastic large cell lymphoma
|
보형물연관 역형성대세포림프종 | dict | 1 | |
| 약물 | Administration-Approved
|
scispacy | 1 | ||
| 약물 | silicone
|
C0037114
silicones
|
scispacy | 1 | |
| 약물 | [BACKGROUND]
|
scispacy | 1 | ||
| 질환 | Cohesive Silicone
|
scispacy | 1 | ||
| 질환 | rupture
|
C3203359
Rupture
|
scispacy | 1 | |
| 질환 | contracture
|
C0009917
Contracture
|
scispacy | 1 | |
| 질환 | breast implant-associated anaplastic large cell lymphoma
|
C4528210
Breast implant-associated anaplastic large-cell lymphoma
|
scispacy | 1 | |
| 질환 | Breast Implants
|
scispacy | 1 | ||
| 질환 | Baker grade III/IV capsular contracture
|
scispacy | 1 | ||
| 기타 | patients
|
scispacy | 1 | ||
| 기타 | patient
|
scispacy | 1 |
MeSH Terms
Adult; Breast; Breast Implantation; Breast Implants; Female; Humans; Implant Capsular Contracture; Lymphoma, Large-Cell, Anaplastic; Middle Aged; Patient Satisfaction; Postoperative Complications; Prospective Studies; Prosthesis Design; Reoperation; Silicone Gels; Surface Properties; United States; United States Food and Drug Administration
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- The Plastic Surgery In-Service Examination: A Scoping Review.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.