CD30 Lateral Flow and Enzyme-Linked Immunosorbent Assays for Detection of BIA-ALCL: A Pilot Study.

Cancers 2023 Vol.15(21)

Zeyl VG, Xu H, Khan I, Machan JT, Clemens MW, Hu H, Deva A, Glicksman C, McGuire P, Adams WP, Sieber D, Sinha M, Kadin ME

관련 도메인

Abstract

[INTRODUCTION] Breast Implant-Associated Anaplastic Large Cell Lymphoma (BIA-ALCL) commonly presents as a peri-implant effusion (seroma). CD30 (TNFRSF8) is a consistent marker of tumor cells but also can be expressed by activated lymphocytes in benign seromas. Diagnosis of BIA-ALCL currently includes cytology and detection of CD30 by immunohistochemistry or flow cytometry, but these studies require specialized equipment and pathologists' interpretation. We hypothesized that a CD30 lateral flow assay (LFA) could provide a less costly rapid test for soluble CD30 that eventually could be used by non-specialized personnel for point-of-care diagnosis of BIA-ALCL.

[METHODS] We performed LFA for CD30 and enzyme-linked immunosorbent assay (ELISA) for 15 patients with pathologically confirmed BIA-ALCL and 10 patients with benign seromas. To determine the dynamic range of CD30 detection by LFA, we added recombinant CD30 protein to universal buffer at seven different concentrations ranging from 125 pg/mL to 10,000 pg/mL. We then performed LFA for CD30 on cryopreserved seromas of 10 patients with pathologically confirmed BIA-ALCL and 10 patients with benign seromas.

[RESULTS] Recombinant CD30 protein added to universal buffer produced a distinct test line at concentrations higher than 1000 pg/mL and faint test lines at 250-500 pg/mL. LFA produced a positive test line for all BIA-ALCL seromas undiluted and for 8 of 10 malignant seromas at 1:10 dilution, whereas 3 of 10 benign seromas were positive undiluted but all were negative at 1:10 dilution. Undiluted CD30 LFA had a sensitivity of 100.00%, specificity of 70.00%, positive predictive value of 76.92%, and negative predictive value of 100.00% for BIA-ALCL. When specimens were diluted 1:10, sensitivity was reduced to 80.00% but specificity and positive predictive values increased to 100.00%, while negative predictive value was reduced to 88.33%. When measured by ELISA, CD30 was below 1200 pg/mL in each of six benign seromas, whereas seven BIA-ALCL seromas contained CD30 levels > 2300 pg/mL, in all but one case calculated from dilutions of 1:10 or 1:50.

[CONCLUSIONS] BIA-ALCL seromas can be distinguished from benign seromas by CD30 ELISA and LFA, but LFA requires less time (<20 min) and can be performed without special equipment by non-specialized personnel, suggesting future point-of-care testing for BIA-ALCL may be feasible.

추출된 의학 개체 (NER)

유형영어 표현한국어 / 풀이UMLS CUI출처등장
합병증 bia-alcl 보형물연관 역형성대세포림프종 dict 11
해부 peri-implant scispacy 1
해부 tumor cells scispacy 1
해부 lymphocytes scispacy 1
해부 LFA → lateral flow assay scispacy 1
해부 line scispacy 1
해부 lines scispacy 1
해부 specimens scispacy 1
해부 breast 유방 dict 1
합병증 seromas scispacy 1
합병증 seroma 장액종 dict 1
합병증 anaplastic large cell lymphoma 보형물연관 역형성대세포림프종 dict 1
약물 pg/mL. scispacy 1
약물 [INTRODUCTION] Breast Implant-Associated Anaplastic Large Cell Lymphoma scispacy 1
약물 [CONCLUSIONS] BIA-ALCL scispacy 1
질환 Breast Implant-Associated Anaplastic scispacy 1
질환 peri-implant effusion scispacy 1
질환 tumor C0027651
Neoplasms
scispacy 1
질환 seromas C0262627
Seroma
scispacy 1
질환 malignant seromas scispacy 1
질환 benign seromas scispacy 1
질환 BIA-ALCL seromas scispacy 1
기타 CD30 scispacy 1
기타 TNFRSF8 scispacy 1
기타 patients scispacy 1
기타 CD30 LFA scispacy 1

🔗 함께 등장하는 도메인

이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들

관련 논문