The current gold standard breast volumetry technique seems to overestimate fat graft volume retention in the breast: A validation study.
Abstract
[BACKGROUND] MRI is generally considered as the gold standard for measuring breast volume because of its high accuracy of the modality. Many techniques used to measure total breast volume have been validated, but none of these techniques have been validated for their ability to measure the volume retention of fat grafts in the breast. In this study, the authors investigated the accuracy of the most common MRI technique used to measure fat graft retention in the breast by measuring the volume changes after breast augmentation.
[METHODS] Patients undergoing breast augmentation with either breast implants or fat grafting underwent MRI scans before and after surgery. Blinded observers measured the change in breast volume from the MRI scans. The difference between the measured change in breast volume and the volume of the breast augmentation was used to determine the accuracy of the MRI technique.
[RESULTS] Twenty-eight patients with a total of 56 breasts were included. In total, 168 measurements of change in breast volume were performed by the observers. The MRI measurements of change in breast volume overestimated the true volumes of the breast augmentations by an average of 50.8%, and only 8 of the 168 individual measurements had measurement errors below 50 mL.
[CONCLUSION] The MRI technique, which is considered as the gold standard for the quantification of fat graft volume retention, was associated with a significant measurement error. These findings have potential implications for the interpretation of previously published results of studies based on this technique.
[METHODS] Patients undergoing breast augmentation with either breast implants or fat grafting underwent MRI scans before and after surgery. Blinded observers measured the change in breast volume from the MRI scans. The difference between the measured change in breast volume and the volume of the breast augmentation was used to determine the accuracy of the MRI technique.
[RESULTS] Twenty-eight patients with a total of 56 breasts were included. In total, 168 measurements of change in breast volume were performed by the observers. The MRI measurements of change in breast volume overestimated the true volumes of the breast augmentations by an average of 50.8%, and only 8 of the 168 individual measurements had measurement errors below 50 mL.
[CONCLUSION] The MRI technique, which is considered as the gold standard for the quantification of fat graft volume retention, was associated with a significant measurement error. These findings have potential implications for the interpretation of previously published results of studies based on this technique.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 15 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 3 | |
| 해부 | fat graft
|
scispacy | 1 | ||
| 해부 | fat grafts
|
scispacy | 1 | ||
| 해부 | fat
|
scispacy | 1 | ||
| 해부 | breasts
|
scispacy | 1 | ||
| 약물 | [BACKGROUND] MRI
|
scispacy | 1 | ||
| 질환 | volume retention
|
scispacy | 1 | ||
| 기타 | Patients
|
scispacy | 1 |
MeSH Terms
Adipose Tissue; Adult; Breast; Breast Implantation; Breast Implants; Female; Humans; Magnetic Resonance Imaging; Mammaplasty; Observer Variation; Organ Size; Young Adult
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.