Characterisation of breast implant surfaces and correlation with fibroblast adhesion.
Abstract
[INTRODUCTION] Capsular contracture formation is a common complication following breast augmentation surgery. Breast implant shells have either a smooth or a textured surface. Smooth surfaces demonstrate a higher incidence of contracture formation. The 3-dimensional surface of textured implants is thought to disrupt contractile forces and reduce capsular contracture rates.
[AIM] To investigate the interaction of fibroblasts with silicone breast implant surfaces through characterization of their unique features.
[METHOD] Surfaces of smooth and textured breast implants were characterized using a confocal laser scanning microscope, a microtest 5 kN tensile testing device, and a contact angle goniometer. The kinetics of fibroblast interaction with these surfaces was further analysed.
[RESULTS] The textured surfaces were rough, and nodular containing high peaks and deep crevasses with roughness (Sa) values in the range 8.88-18.83 μm and contact angles between 130° and 142°. The smooth implant surfaces were less rough, more regular and repetitive with 0.06-0.07 μm surface roughness, and contact angles between 110.9° and 111.8°. The textured surfaces displayed higher bending stiffness than the smooth surfaces (0.19 and 0.26 N mm). Significant (p<0.05) numbers of fibroblasts were attached to the textured surfaces compared to the smooth surfaces which had higher levels of cell adhesion with surface roughness above 8 μm and contact angles above 130°.
[CONCLUSIONS] In summary, surfaces with arithmetical mean deviation of greater roughness and reduced hydrophilicity with high water contact angles enhanced cell adhesion. These features aid design of improved surfaces, which may help, in prevention of breast capsular formation.
[AIM] To investigate the interaction of fibroblasts with silicone breast implant surfaces through characterization of their unique features.
[METHOD] Surfaces of smooth and textured breast implants were characterized using a confocal laser scanning microscope, a microtest 5 kN tensile testing device, and a contact angle goniometer. The kinetics of fibroblast interaction with these surfaces was further analysed.
[RESULTS] The textured surfaces were rough, and nodular containing high peaks and deep crevasses with roughness (Sa) values in the range 8.88-18.83 μm and contact angles between 130° and 142°. The smooth implant surfaces were less rough, more regular and repetitive with 0.06-0.07 μm surface roughness, and contact angles between 110.9° and 111.8°. The textured surfaces displayed higher bending stiffness than the smooth surfaces (0.19 and 0.26 N mm). Significant (p<0.05) numbers of fibroblasts were attached to the textured surfaces compared to the smooth surfaces which had higher levels of cell adhesion with surface roughness above 8 μm and contact angles above 130°.
[CONCLUSIONS] In summary, surfaces with arithmetical mean deviation of greater roughness and reduced hydrophilicity with high water contact angles enhanced cell adhesion. These features aid design of improved surfaces, which may help, in prevention of breast capsular formation.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 해부 | breast
|
유방 | dict | 6 | |
| 합병증 | capsular contracture
|
피막구축 | dict | 2 | |
| 시술 | breast augmentation
|
유방성형술 | dict | 1 | |
| 해부 | fibroblast
|
scispacy | 1 | ||
| 해부 | smooth
|
scispacy | 1 | ||
| 해부 | contractile
|
scispacy | 1 | ||
| 해부 | fibroblasts
|
scispacy | 1 | ||
| 해부 | surface
|
scispacy | 1 | ||
| 해부 | cell
|
scispacy | 1 | ||
| 합병증 | breast implant
|
scispacy | 1 | ||
| 합병증 | nodular
|
scispacy | 1 | ||
| 재료 | silicone breast implant
|
실리콘 보형물 | dict | 1 | |
| 약물 | [INTRODUCTION] Capsular contracture
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | contracture
|
C0009917
Contracture
|
scispacy | 1 | |
| 질환 | bending stiffness
|
scispacy | 1 | ||
| 질환 | Breast implant shells
|
scispacy | 1 | ||
| 기타 | breast capsular
|
scispacy | 1 |
MeSH Terms
Biocompatible Materials; Breast Implants; Cell Adhesion; Cells, Cultured; Elastic Modulus; Equipment Failure Analysis; Fibroblasts; Humans; Materials Testing; Prosthesis Design; Silicon; Statistics as Topic; Surface Properties; Tensile Strength
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- The impact of three-dimensional simulation and virtual reality technologies on surgical decision-making and postoperative satisfaction in aesthetic surgery: a preliminary study.
- Cutaneous fistula of the breast: A complication of cosmetic autologous fat transfer.
- Epidermal inclusion cyst after breast reduction mammoplasty.
- Clinical outcomes of synthetic absorbable mesh use in breast surgery: First case series in reconstruction and aesthetic mastopexy.
- Implant-based versus autologous mastopexy after massive weight loss: Complications and patient satisfaction.