Robot-Assisted Microsurgery Has a Steeper Learning Curve in Microsurgical Novices.
Abstract
[INTRODUCTION] Mastering microsurgery requires advanced fine motor skills, hand-eye coordination, and precision, making it challenging for novices. Robot-assisted microsurgery offers benefits, such as eliminating physiological tremors and enhancing precision through motion scaling, which may potentially make learning microsurgical skills easier.
[MATERIALS AND METHODS] Sixteen medical students without prior microsurgical experience performed 160 anastomoses in a synthetic model. The students were randomly assigned into two cohorts, one starting with the conventional technique (HR group) and one with robotic assistance (RH group) using the Symani surgical system.
[RESULTS] Both cohorts showed a reduction in procedural time and improvement in SAMS scores over successive attempts, with robotic anastomoses demonstrating a 48.2% decrease in time and a 54.6% increase in SAMS scores. The decreases were significantly larger than the RH group ( < 0.05). The quality of the final anastomoses was comparable in both groups ( > 0.05).
[DISCUSSION] This study demonstrated a steep preclinical learning curve for robot-assisted microsurgery (RAMS) among novices in a synthetic, preclinical model. No significant differences in SAMS scores between robotic and manual techniques after ten anastomoses. Robot-assisted microsurgery required more time per anastomosis, but the results suggest that experience with RAMS may aid in manual skill acquisition. The study indicates that further exploration into the sequencing of robotic and manual training could be valuable, especially in designing structured microsurgical curricula.
[MATERIALS AND METHODS] Sixteen medical students without prior microsurgical experience performed 160 anastomoses in a synthetic model. The students were randomly assigned into two cohorts, one starting with the conventional technique (HR group) and one with robotic assistance (RH group) using the Symani surgical system.
[RESULTS] Both cohorts showed a reduction in procedural time and improvement in SAMS scores over successive attempts, with robotic anastomoses demonstrating a 48.2% decrease in time and a 54.6% increase in SAMS scores. The decreases were significantly larger than the RH group ( < 0.05). The quality of the final anastomoses was comparable in both groups ( > 0.05).
[DISCUSSION] This study demonstrated a steep preclinical learning curve for robot-assisted microsurgery (RAMS) among novices in a synthetic, preclinical model. No significant differences in SAMS scores between robotic and manual techniques after ten anastomoses. Robot-assisted microsurgery required more time per anastomosis, but the results suggest that experience with RAMS may aid in manual skill acquisition. The study indicates that further exploration into the sequencing of robotic and manual training could be valuable, especially in designing structured microsurgical curricula.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | microsurgery
|
미세수술 | dict | 5 | |
| 기법 | robot-assisted
|
로봇수술 | dict | 4 | |
| 약물 | [INTRODUCTION] Mastering microsurgery
|
scispacy | 1 | ||
| 질환 | tremors
|
C0040822
Tremor
|
scispacy | 1 | |
| 기타 | SAMS
|
scispacy | 1 |
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.