New Robotic System with Wristed Microinstruments Allows Precise Reconstructive Microsurgery: Preclinical Study.
Abstract
[BACKGROUND] Microsurgery allows complex reconstruction of tissue defects after oncological resections or severe trauma. Performing these procedures may be limited by human tremor, precision, and manual dexterity. A new robot designed specifically for microsurgery with wristed microinstruments and motion scaling may reduce human tremor and thus enhance precision. This randomized controlled preclinical trial investigated whether this new robotic system can successfully perform microsurgical needle driving, suturing, and anastomosis.
[METHODS] Expert microsurgeons and novices completed six needle passage exercises and performed six anastomoses by hand and six with the new robot. Experienced microsurgeons blindly assessed the quality of the procedures. Precision in microneedle driving and stitch placement was assessed by calculating suturing distances and angulation. Performance of microsurgical anastomoses was assessed by time, learning curves, and the Anastomosis Lapse Index score for objective performance assessment.
[RESULTS] Refined precision in suturing was achieved with the robot when compared with the manual technique regarding suture distances (p = 0.02) and angulation (p < 0.01). The time required to perform microsurgical anastomoses was longer with the robot, however, both expert and novice microsurgeons reduced times with practice. The objective evaluation of the anastomoses performed by novices showed better results with the robot.
[CONCLUSIONS] This study demonstrated the feasibility of performing precise microsutures and anastomoses using a new robotic system. Compared to standard manual techniques, robotic procedures were longer in time, but showed greater precision.
[METHODS] Expert microsurgeons and novices completed six needle passage exercises and performed six anastomoses by hand and six with the new robot. Experienced microsurgeons blindly assessed the quality of the procedures. Precision in microneedle driving and stitch placement was assessed by calculating suturing distances and angulation. Performance of microsurgical anastomoses was assessed by time, learning curves, and the Anastomosis Lapse Index score for objective performance assessment.
[RESULTS] Refined precision in suturing was achieved with the robot when compared with the manual technique regarding suture distances (p = 0.02) and angulation (p < 0.01). The time required to perform microsurgical anastomoses was longer with the robot, however, both expert and novice microsurgeons reduced times with practice. The objective evaluation of the anastomoses performed by novices showed better results with the robot.
[CONCLUSIONS] This study demonstrated the feasibility of performing precise microsutures and anastomoses using a new robotic system. Compared to standard manual techniques, robotic procedures were longer in time, but showed greater precision.
추출된 의학 개체 (NER)
| 유형 | 영어 표현 | 한국어 / 풀이 | UMLS CUI | 출처 | 등장 |
|---|---|---|---|---|---|
| 시술 | microsurgery
|
미세수술 | dict | 3 | |
| 해부 | tissue
|
scispacy | 1 | ||
| 약물 | [BACKGROUND] Microsurgery
|
scispacy | 1 | ||
| 약물 | [CONCLUSIONS]
|
scispacy | 1 | ||
| 질환 | trauma
|
C0043251
Wounds and Injuries
|
scispacy | 1 | |
| 질환 | tremor
|
C0040822
Tremor
|
scispacy | 1 | |
| 질환 | angulation
|
C0333179
Angulated
|
scispacy | 1 | |
| 기타 | human
|
scispacy | 1 |
MeSH Terms
Anastomosis, Surgical; Humans; Microsurgery; Robotic Surgical Procedures; Robotics; Suture Techniques; Tremor
📑 인용 관계
🔗 함께 등장하는 도메인
이 논문이 속한 카테고리와 같은 논문에서 자주 함께 다뤄지는 카테고리들
관련 논문
- Endodontic implications of hypercementosis: A systematic review of anatomical challenges and therapeutic strategies.
- Breast plastic surgery in perimenopausal and postmenopausal women: Menopause-informed counseling on screening, safety, and long-term breast health.
- Application of the SCIA-Pure Skin Perforator Flap in Bilateral Upper Eyelid Reconstruction: A Case Report and Review of the Literature.
- Free flap reconstruction of a cast-related pressure ulcer in a pediatric patient with spinal muscular atrophy.
- Characterization of Trimmed Nerve Morphology Using High-Resolution Imaging: Comparison of Three Surgical Instruments.